Wellness for your instruments: Discover MELAcare now>

|
    • Unbenannt-1Best Sellers
    • Autoclave Tile
      • premium-klassePremium Class
      • Prime LinePrime Line (Class B)
      • Pro Line Product picturePro Line (Class B)
      • MicrosoftTeams-image (1196)-minPro Line (Class S)
      • 150 2Vacuclave 105 / 305
      • Vacuclave 550 Menu ElementVacuclave 550
      • Grossautoklaven-SystemloesungCliniclave 45 Series
      • 24-BL-StoererVacuklav 24 BL+
      • MELAtronic ProduktfotoMELAtronic Series
      • Zubehör Careclave ProduktfotoAccessories & components for autoclaves
    • Grossautoklaven-Systemloesung
      • Cliniclave 45 Produktfoto für Kaegorie MenuCliniclave 45
      • Cliniclave 45 M ProduktfotoCliniclave 45 M
      • Cliniclave 45 D ProduktfotoCliniclave 45 D
      • Cliniclave 45 MD ProduktfotoCliniclave 45 MD
      • Vacuclave 550 Menu ElementVacuclave 550
      • Großautoklaven Zubehör ProduktfotoAccessories & components for large autoclaves
    • 150 2
      • 150-ohne 1Vacuclave 105
      • 350-ohne 1Vacuclave 305
      • MELAquick ProduktfotoMELAquick Series
      • MELAtronic ProduktfotoMELAtronic Series
    • mt20-gross
      • mt20-grossMELAtherm 20
      • MT10EVO_Produkt_geschlossenMELAtherm 10 Evolution
      • Körbe und Adapter ProduktfotoLoading patterns
      • Collage-Korb 3-minBaskets & Adapters
      • MEtherm Process Agents ProduktfotoMEtherm Process Agents
      • teaser-washcheckMELAcontrol Wash Check
      • Regenerating SaltRegenerating Salt
      • MELAtherm Zubehör MenuAccessories & components for thermal disinfectors
    • MELAcare 1
      • MELAcare 1 (2) MELAcare
      • MELAcare-Box-BlueMELAcare Lid Blue
      • MELAcare-Box-GreenMELAcare Lid Green
      • Care Oil ProduktfotoCare Oil
      • melacare Oil Lid adapter 02Accessories & components for MELAcare
    • Careclave_Produkt_offen für Menü
      • Careclave_Produkt_offen für MenüCareclave
      • Menu Foto Carebox BlueCarebox Blue
      • Carebox Green Menu BildCarebox Green
      • Cooling Box ProduktfotoCooling Box
      • Care Oil ProduktfotoCare Oil
      • Zubehör Careclave ProduktfotoAccessories & components for Careclave
    • MicrosoftTeams-image (1170)-min
      • MicrosoftTeams-image (1420)-min(1)MELAseal 300
      • MELAseal 200 Menu FotoMELAseal 200
      • MELAseal 100 Bild MenuMELAseal 100+
      • MELAfol Menu BildMELAfol+
      • MELAcontrol Seal Check Routine Prüfung MenuRoutine Testing
      • Siegelnahtfestigkeit MaschineSeal Seam Stability Test
      • Zubehör MELAseal MenuAccessories & components for sealing devices
    • MicrosoftTeams-image (1169)-min
      • Trays neuMELAstore Trays
      • OTS MenüMELAstore Tray Ophthalmology
      • MELAstoreMELAstore Boxes
      • SterilisierbehälterSterilization Containers
      • me01100 1MELAstore transport box
      • MELAfol Menu BildMELAfol+
      •   SterilisationspapierSterilisation paper
      • Zubehör PackagingAccessories & components for sterile goods storage
    • MicrosoftTeams-image (1168)-min
      • MELAtrace-Pro Menu IconMELAtrace
      • MELAverse Produkt KachelMELAverse
      • MELAprint 80MELAprint 80
      • MELAprint 60 ProduktfotoMELAprint 60
      • MELAdoc Titelbild-blau MELAdoc
      • MELAcontrol Menu FotoMELAcontrol
      • MELAflash ProduktfotoMELAflash
      • MELAnetBox Menu BildMELAnet Box
      • Zubehör DokuAccessories & components for documentation
    • MELAverse Produkt KachelMELAverse
    • MicrosoftTeams-image (1171)
      • MicrosoftTeams-image (1052)Care Oil Spray
      • MicrosoftTeams-image (1053)Chamber Protect
    • MicrosoftTeams-image (1172)-min
      • MELAdem 40 ProduktfotoMELAdem 40
      • MELAdem 47 ProduktfotoMELAdem 47
      • MELAdem 53 Menu FotoMELAdem 53
      • MELAdem 53 CMELAdem 53 C
      • MELAdem 56 ProduktfotoMELAdem 56
      • MELAjet ProduktfotoMELAjet
      • Zubehör WasserAccessories & components for water treatment
    • MicrosoftTeams-image (1167)-min
      • MELAcontrol Menu FotoMELAcontrol Helix
      • MELAcontrol ProMELAcontrol Pro
      • MELAcontrol Type 5MELAcontrol Type 5
      • MELAcontrol Bowie & Dick TestMELAcontrol Bowie & Dick Test
      • teaser-washcheckMELAcontrol Wash Check
      • MELAcontrol Seal Check Routine Prüfung MenuMELAcontrol Seal Check
      • Siegelnahtfestigkeit MaschineSeal Seam Stability Test
      • Menu ink TestMELAcontrol Ink Test
    • MELAfol ProduktfotoConsumables
    • Downloadcenter Tile ImageDownloadcenter
    • Service-troubleshootingTroubleshooting
    • Video Tutorials Autoclave
      • Balkenteaser-MELAcare 1MELAcare Tutorials
      • Careclave Tutorial MenuCareclave Tutorials
      • Autoklaven Tutorial Autoclaves Tutorials
      • Tutorial Thermodesinfektor MenuWasher-Disinfector Tutorials
      • Siegelgeräte Menu TutorialMELAseal 200 Tutorials
      • MS300 BalkenteaserMELAseal 300 Tutorials
      • MELAtrace TutorialMELAtrace Tutorials
    • Gerätertegistrierung Menu BildDevice Registration
    • Techniker-schulung Menü BildTechnical Training
    • KonfiguratorConfigurator
    • save money Promotions
    • Entdecken-referenzSuccess Stories
    • Hygiene Hub TileInspiration & Tips
    • Tutorials TileNews
    • MELAG FamilieFamily Business & Jobs
    • Beratung-uebersicht
      • Beratung-uebersichtBook a consultation: On-site, online or in a showroom
      • entdecken-ansprechpartnerYour sales contacts
      • Messen & Kongresse Bild MenuFairs & Congresses
      • beratung-kontaktformularContact form
      • Gebauer GFCompany & Apprenticeship
    • entdecken-ansprechpartner
      • Beratung-uebersichtBook a consultation: On-site, online or in a showroom
      • entdecken-ansprechpartnerYour sales contacts
      • Messen & Kongresse Bild MenuFairs & Congresses
      • beratung-kontaktformularContact form
      • Gebauer GFCompany & Apprenticeship
    • Messen & Kongresse Bild MenuFairs & Congresses
    • beratung-kontaktformularContact form
    • Gebauer GFCompany & Apprenticeship
    • Dental
    • Medical
    • Ophthalmo
    • Hospital
    • Veterinary
    • Podiatry
    • University
Select language and country
Select language and country

Newsletter Subscription

Register for our newsletter and always stay up to date.

Products
  • Instrument Reprocessing
  • Autoclaves
  • Large Autoclaves
  • Washer-Disinfectors
  • MELAcare
  • Careclave
  • Sealing Devices
Service
  • Downloads
  • Troubleshooting
  • Video Tutorials
  • Blog
Company
  • We are MELAG
  • Sustainability
  • History
  • News
  • Contact
  • Instrument Reprocessing
  • Autoclaves
  • Large Autoclaves
  • Washer-Disinfectors
  • MELAcare
  • Careclave
  • Sealing Devices
  • Downloads
  • Troubleshooting
  • Video Tutorials
  • Blog
  • We are MELAG
  • Sustainability
  • History
  • News
  • Contact
  • Imprint
  • Terms and Conditions
  • Privacy
  • Whistleblower
  • Open Cookie Settings

© MELAG Medizintechnik GmbH & Co. KG

Placeholder for alt text
Placeholder for alt text
Placeholder for alt text
Placeholder for alt text
Placeholder for alt text
Placeholder for alt text
Placeholder for alt text
Placeholder for alt text
Placeholder for alt text
Placeholder for alt text
Placeholder for alt text
Placeholder for alt text
Placeholder for alt text

MELAstore Tray 100

Order no. ME01178

Specially developed for the MELAtherm® 10 and MELAthem® 20 Series, the MELAstore® Tray 100 (dimensions BxHxT 17,9 x 3 x 28  cm) made of high-quality stainless steel ensures excellent cleaning results due to its large openings and enables quick and safe handling of individual instruments and instrument sets (e.g., for periodontal treatments, conservative treatments or minor surgical procedures). The advantages include a significant time saving, increased processing quality through standardisation, effective damage prevention as well as improved safety for the practice team. Variable silicone supports ensure a stable and secure fixation of the instruments. The ability to label each tray with an individual identification label supports seamless documentation of the reprocessing procedures.

Do you have questions?

Then use our live consulting service via chat or telephone:

+49 30 75 79 11-0info@melag.de

Request consultation

Please send us your individual product configuration using the registration form.

MELAstore® Tray 100

Maximum efficiency and safety in instrument reprocessing

The robust MELAstore® Tray 100 (width 17.9 cm x height 3 cm x depth 28 cm) made of high-quality stainless steel is ideal for the safe and efficient reprocessing of individual instruments and instrument sets. It is particularly suitable for procedures such as periodontal treatments, conservative treatments and minor surgical procedures.

Designed for use with the MELAtherm® 10 series, this tray ensures optimal cleaning and sterilization of instruments. Each tray can be provided with a label or a colored silicone strip to enable precise allocation and documentation during the entire decontamination cycle.

ME01181
Placeholder for alt text

Enable YouTube content

To play this video, please allow cookies and other external services from YouTube.

MT10-Evo-Beladung-min
Unlock a World of Benefits:

Absolute safety, unsurpassed efficiency.

Discover the ultimate solution to enhance your practice's efficiency and safety with our innovative features. The MELAstore Trays are designed to streamline your instrument processing while safeguarding your team and patients. Experience the advantages of precise customization and robust protection—all tailored to fit your clinical needs perfectly. Dive into a world where every detail is crafted for perfection.

The MELAstore Tray 100 ensures maximum time efficiency: a tray can be loaded and removed more quickly than a large number of individual instruments. Thanks to its intelligent design, this is not only child's play, but also safe, thus reducing injuries to your staff.

Made from high-quality stainless steel, our trays are durable and easy to clean. This not only saves you downtime and replacements, but also provides long-term protection for your patients and your team.

The MELAstore Tray 100 improves workplace safety by safely organizing instruments to prevent sharps injuries, facilitating thorough sterilization to reduce contamination risks, and ensuring long-lasting, reliable protection for both staff and patients.

Each tray can be equipped with a labeling tag, allowing for precise assignment and documentation throughout the processing cycle. Safe and Adjustable: Variable silicone dividers ensure a secure and protective fixation of instruments, further minimizing the risk of damage and injuries.

Tray systems in dental practices: Efficient instrument reprocessing with wash trays

A well-organised set of instruments saves time, reduces the risk of errors and makes it easier to adhere to standardised hygiene procedures. This is precisely where a tray system comes into its own in a dental practice: instruments are no longer moved around the practice on a day-to-day basis individually and in varying combinations, but are prepared, used, reprocessed and made available again as clearly defined sets.

Wash-trays can support the entire instrument cycle – from treatment through mechanical cleaning and disinfection to packaging, sterilisation and storage. However, it is crucial that the trays, instruments, cleaning and disinfection equipment, sterilisation containers and autoclave are coordinated with one another and integrated into the practice’s validated processes.

What is an instrument wash-tray system?

A wash tray is a reusable instrument holder in which the instruments required for a specific treatment are systematically arranged and secured. Depending on the design, flexible silicone ridges or other retainers hold the instruments in a defined position.

The wash tray accompanies the instrument set through several stages of reprocessing. It can – provided the manufacturer has designed it for this purpose – be used to hold the instruments during mechanical cleaning and disinfection and subsequently within a suitable sterilisation container.

However, a wash tray is not automatically a sterile packaging system. Only a sterile barrier system designed and validated for this purpose – for example, a compatible sterilisation container with a suitable filter – enables packaged sterilisation and the protection of sterile items during transport and storage. The requirements for sterile barrier systems and packaging systems are set out in DIN EN ISO 11607-1.

Dental tray, wash tray and sterilisation container: what is the difference?

These terms are sometimes used interchangeably in everyday practice, but they refer to different functions.

A dental tray is, first and foremost, an instrument holder designed to organise the layout of instruments during treatment. Not every dental tray is automatically suitable for machine cleaning, thermal disinfection or steam sterilisation.

A wash tray is designed for reprocessing in a cleaning and disinfection machine. Its design is intended to ensure that water, process chemicals and the rinsing mechanism can adequately reach the instruments and that liquids can subsequently drain away.

A sterilisation container, together with its closure and filter system, forms the packaging for sterilisation. When used as intended, the appropriate sterile barrier system protects the sterilised instruments from recontamination until they are used.

This creates a coordinated system: the instrument set remains organised in the wash tray, whilst the container acts as the sterile packaging.

Why is a tray system worthwhile in a dental practice?

Instrument reprocessing involves many repetitive steps. If instruments are sorted, transported, loaded into the washer-disinfector, checked and packed individually, each batch requires a considerable amount of manual effort.

A standardised tray system can simplify and standardise these processes

1. Less time required

A fully assembled instrument set is immediately available before treatment. The practice team does not have to search for the required instruments individually from different drawers or containers each time.

Even after treatment, the instruments remain organised as a set. This reduces the time spent on sorting, loading and packaging. Coordinated systems comprising wash trays and sterilisation containers are designed to guide instrument sets as seamlessly as possible through the entire process, from use through reprocessing to storage.

2. Standardised treatment sets

Fixed sets can be defined for common procedures, for example for:

  • prophylaxis,

  • conservative treatment,

  • periodontology,

  • endodontics,

  • extractions,

  • implantology,

  • oral surgery procedures.

The standardised composition simplifies preparation and makes it immediately clear if an instrument is missing. New staff members also find it easier to follow the established set structures.

3. Improved workplace safety

Pointed and sharp instruments pose a risk of injury during transport, sorting and reprocessing. If they are secured in suitable holders, they need to be picked up and moved individually less often.

However, the tray system is no substitute for safe working practices. Contaminated instruments must still be collected and transported in such a way that staff are not put at risk of puncture or cut injuries or exposure to biological agents. A suitable, closed transport container must therefore be provided for the journey from the treatment room to the reprocessing room. TRBA 250 sets out the specific occupational health and safety requirements for handling biological agents in the healthcare sector.

4. Protection of instruments

Instruments lying loose in baskets or containers may knock against one another, become entangled or be damaged. A defined method of securing them can better protect delicate tips, cutting edges and surfaces.

It is essential that the holders are suitable for the instruments. Silicone bars that are too tight or incorrectly positioned can cause pressure marks, obscure instruments or hinder cleaning.

5. Clear documentation

Coloured silicone dividers, labels, set numbers or barcodes enable unambiguous allocation. For example, a tray can be assigned to a specific type of treatment, a reprocessing batch or a defined loading pattern.

This supports batch documentation and traceability. However, authorisation remains a professional decision to be made by authorised staff and cannot be replaced by colour coding or a barcode alone.

The instrument cycle with a wash tray

The greatest benefit is achieved when the tray system is planned not as a standalone accessory, but as an integral part of a seamless workflow.

1. Compiling the instrument set

First, it is determined which instruments are actually required for the respective procedure. Unnecessary instruments should not be included in the set simply out of habit, as each instrument must undergo the specified reprocessing procedure after use or contamination.

The instruments are arranged in the wash tray in such a way that they are clearly identifiable and securely held in place. Articulated instruments must be opened, dismantled or positioned in a manner suitable for cleaning, in accordance with the manufacturer’s instructions.

The configuration should be documented in an illustrated packing or tray list. This enables the team to quickly check that everything is present and to reproduce the set-up at any time.

2. Preparation in the treatment room

The reprocessed set, which may be sterile-packed, is only opened immediately before treatment. The instruments remain clearly arranged and can be removed in a specified order.

A well-organised tray therefore not only supports hygiene but also assists with the treatment. Missing or incorrectly sorted instruments are noticed at an early stage.

3. Safe collection and transport

After treatment, the used instruments are returned to the tray, provided this can be done without posing any additional risk and is part of the established workflow. Pointed and sharp ends should be positioned so that no unnecessary risk of injury arises during further handling.

The contaminated wash tray is then transported to the reprocessing room in a closed, sufficiently robust and clearly labelled container. Transporting contaminated instruments openly through the practice should be avoided.

4. Cleaning and disinfection in the washer-disinfector

The tray is loaded into the cleaning and disinfection machine in accordance with the validated loading pattern. To ensure effective cleaning, all instrument surfaces must be accessible to water and the process chemicals.

In particular, the following points must be observed:

  • Instruments must not completely cover one another.

  • Spray arms and rinsing dead zones must not be obstructed.

  • Hinges must be sufficiently open.

  • Liquids must be able to drain from the tray and the instruments.

  • The maximum load capacity must not be exceeded.

  • Manufacturer’s specifications for instruments, trays and the washer-disinfector must be consistent.

For Class B medical devices, machine-based cleaning and disinfection is generally the method of choice. The DIN-EN-ISO 15883 series sets out requirements for cleaning and disinfection equipment, their processes, loading carriers, validation and routine monitoring.

Special considerations for hollow-body instruments

Not every instrument can be adequately cleaned internally simply by being placed in a wash tray. Instruments with lumens, transmission instruments, suction cannulas or other internal areas may require suitable connectors, adaptors or special holding devices.

Whether an instrument must be connected to an injector system within the tray or separately is determined by the reprocessing instructions provided by the instrument and equipment manufacturer. The external arrangement within the tray must not result in necessary internal cleaning being omitted.

5. Visual inspection, maintenance and functional testing

Following the automated process, the tray is removed in the clean area. Even if an RDG programme has been successfully completed, the instruments must still be inspected.

The following are checked in particular:

  • cleanliness,

  • dryness,

  • Integrity,

  • corrosion or surface changes,

  • mobility of joints,

  • functionality,

  • completeness of the set.

Instruments that have not been adequately cleaned must not be released for sterilisation. They must undergo the specified cleaning and disinfection process again. Damaged instruments or those that can no longer be safely reprocessed are discarded.

Maintenance procedures must be carried out in accordance with the manufacturer’s instructions. Only suitable agents that do not impair the subsequent sterilisation process may be used.

6. Packing in the sterilisation container

If the instruments are to be used in a sterile condition, the inspected wash tray is placed in a suitable sterilisation container. The filter system, seal, closure and, where applicable, the tamper-evident seal must be complete and in working order.

Before each use, the following should be checked, amongst other things:

  • Is the container clean and undamaged?

  • Are the lid and base dimensionally stable?

  • Is the seal intact?

  • Is the filter correctly fitted and suitable for the intended use?

  • Do the fastenings and locks work properly?

  • Is the labelling clear?

  • Does the tray fit into the container without tilting or being overloaded?

Only a packaging system that has been assembled in accordance with its intended use and is suitable for the process can provide the intended sterile barrier.

7. Sterilisation in the autoclave

The closed container is loaded into the autoclave in accordance with the validated loading pattern. The specifications regarding positioning, stackability, load quantity and programme selection must be adhered to.

The sterilisation programme must be suitable for the instruments, the container and the load. The daily routine load must be covered by the validation or performance qualification of the sterilisation process.

The introduction of an additional tray system alters the loading configuration. Therefore, new tray or container configurations must not be incorporated into routine operations without prior assessment. Where necessary, the loading patterns may need to be amended or a new performance assessment carried out.

8. Release and Documentation

Following the sterilisation process, the process sequence, relevant parameters and required batch controls are evaluated. The container is then checked for dryness, integrity, correct sealing and complete labelling.

Release must be carried out by an authorised and qualified person. They should be able to clearly identify at least the following information:

  • steriliser and batch number,

  • programme used,

  • date of reprocessing,

  • tray or instrument set,

  • result of the process control,

  • Release decision,

  • person authorising.

A successfully completed device programme alone does not replace the documented batch and sterile goods release.

9. Storage and re-issuing

Approved containers are stored under suitable conditions, protected from moisture, contamination and mechanical damage. The storage system should follow the ‘first in, first out’ principle and allow for clear stock control.

Before the next treatment, the container is checked for integrity and taken to the treatment room. Only there is it opened under aseptic conditions and the tray prepared.

If the container is damaged, damp, has been opened unintentionally or is not clearly labelled, its contents must not be used as sterile.

Which instruments are suitable for a tray system?

Tray systems are particularly suitable for regularly recurring combinations of instruments. These include, for example, basic instrument sets, prophylaxis sets and instrument sets for endodontics, periodontology, surgery or implantology.

However, suitability does not depend solely on size. Before including an instrument, the following questions must be clarified:

  • Is it suitable for machine cleaning and thermal disinfection?

  • Can it be reprocessed whilst fixed in the tray?

  • Do any joints need to be opened or parts dismantled?

  • Does it require internal cleaning via an adapter?

  • Is it compatible with the process chemicals used?

  • Is it suitable for steam sterilisation?

  • Is it covered by the validated process?

The manufacturer’s specifications for the medical device take precedence. A tray system can standardise reprocessing, but cannot compensate for an instrument’s lack of reprocessability.

Practices should bear the following in mind when making their selection

A suitable wash tray should not be selected solely on the basis of size or purchase price. The key factor is how well it interacts with existing equipment and processes.

  • Compatibility: The tray, holders and containers must be approved for or suitable for use with the existing cleaning and disinfection equipment and the autoclave. The usable capacities and loading racks must also be taken into account.

  • Easy-to-clean design: The tray should feature open surfaces, good water flow and reliable drainage. Inaccessible gaps and areas where residual water or dirt can accumulate are undesirable.

  • Flexible instrument holders: Silicone bars should be configurable so that different instrument sets can be securely held in place. At the same time, they must not unnecessarily cover the surfaces to be cleaned.

  • Clear labelling: Colour codes , labels or barcodes facilitate allocation to treatment, set, batch and storage location. However, colour coding should not be the sole form of identification, as colours can be confused or interpreted differently by individual staff members.

  • Ergonomics and health and safety: The tray should be easy to grip and transport safely. Sharp instruments must not protrude unprotected from the tray. The weight of a fully loaded tray must also be taken into account in day-to-day use.

  • Availability of suitable containers: If the instrument setis to be packed, sterilised and stored, a compatible sterilisation container must be available. The entire system should be able to be integrated into the practice workflow without the need for makeshift combinations.

Common mistakes when using wash trays

A tray system only improves processes if it is planned and used correctly. Typical errors include:

  • trays packed too densely,

  • incorrectly positioned or closed articulated instruments,

  • hollow instruments not properly secured,

  • holders that cover large areas of the instrument surfaces,

  • mixing of incomplete treatment sets,

  • failure to carry out visual and functional checks following the RDG process,

  • use of the wash tray without a suitable sterile barrier system,

  • damaged or incorrectly fitted container filters,

  • undocumented changes to tray loading,

  • use of new sets outside the validated load configurations.

Particularly problematic is the assumption that a properly arranged tray automatically guarantees a hygienically safe result. The demonstrable effectiveness of the entire validated reprocessing process remains crucial.

How a practice successfully introduces a tray system

The transition should take place in stages. First, the most frequently used instrument combinations are analysed and grouped into practical sets.

The following procedure is then recommended:

  • Record the instrument inventory and treatment procedures.

  • Define suitable instrument sets.

  • Check compatibility with the instrument washer-disinfector, autoclave and containers.

  • Determine tray configurations and document them photographically.

  • Update work instructions and loading patterns.

  • Assess the impact on validated processes.

  • Provide practical training for staff.

  • Test the system initially with selected sets.

  • Evaluate experience gained from treatment and reprocessing.

  • Optimise the tray configurations where necessary.

In doing so, it is not only the maximum instrument capacity that should be considered. The crucial factor is whether the number of trays is appropriate for the instrument stock, patient volume and actual reprocessing cycles.

Conclusion: A systematic approach to the entire instrument cycle

A tray system for a dental practice can do far more than simply provide a tidy place to store instruments. When properly planned, it supports the entire workflow – from treatment through cleaning, disinfection and sterilisation to storage and reissuing.

The key benefits lie in standardisation: instrument sets are clearly structured, need to be rearranged individually less often, and can be more easily monitored, documented and allocated. This saves valuable working time whilst also providing better protection for both instruments and the practice team.

However, hygienic success does not depend on the tray alone. It is based on the coordinated interaction of suitable instruments, correct tray configuration, safe transport, validated machine cleaning and disinfection, careful inspection, a suitable sterile barrier system, validated sterilisation and documented release.

A good wash tray therefore does more than just organise instruments – it structures the entire reprocessing process.

NEW

Multicolored holders for MELAstore trays

The MELAstore trays enable structured storage of the instruments. 10 different colored silicone bars enable quick identification of the individual sets. This saves time, avoids mix-ups and helps you to consistently implement high hygiene standards, tailored to your treatment focus or practice.

Learn More
stege-melastore 100
MELAstore Amortisation

An investment in MELAstore amortizes itself within a year.

The use of only a single MELAstore Tray 100 and a MELAstore Box 100 for 14 instruments saves 8 minutes in comparison to separate sealing with a sterilization packaging. Overall, MELAstore saves you up to 24 minutes per reprocessing process and saves the team valuable time for other tasks.

Using MELAstore just once a day means that the purchase of the Trays and Boxes will be amortized in less than a year. The systematic use of the MELAstore system multiple times in a single day shortens the payback period significantly.

Arrange consultation